1030S Solids Analysis Module

Ürünleri görüntüle
Ürünleri görüntüle

Need to analyze total organic carbon (TOC) or total carbon (TC) in solid materials? That’s exactly why we built the 1030S Solids Analysis Module. This sample preparation module is the perfect complement to the Aurora 1030W or 1030C TOC analyzers.

 

  • Overview

     

    Smartly designed, the compact instrument saves bench space while it is hard at work combusting solid materials, soils, sludge, slurries, and powders for TC and TOC analysis.

    Principle of Operation

    Total organic carbon (TOC) is determined by manually transferring and weighing a solid sample into a quartz sample cup. The total inorganic carbon (TIC) content is removed from the sample by adding acid and heating to drive off CO2 released from inorganic carbon compounds. In the case of total carbon (TC) analysis, the pre-acidification step to remove TIC content is not required.

    The sample cup with the TIC-free sample is placed on the lift mechanism and raised into the combustion tube of the 1030S. The sample is heated to 500° – 900 °C inside the furnace. Organic matter in the sample is oxidized and converted to CO2 which is collected in a one liter capacity gas sampling bag. When the combustion cycle is complete an aliquot of the CO2 sample gas is transferred to the NDIR detector in the Aurora 1030 analyzer for measurement of the mass of carbon in the sample.

    1030S Features:

    • Combusts solid materials soils, sludge, slurries, and powders for TC and TOC analysis
    • Interfaces with Aurora 1030 TOC analyzer allowing analysis of both solid and liquid samples using a single instrument
    • Gas sampling device (patent pending) collects carbon dioxide (CO2) from sample combustion for quantitation and analysis of intra-sample replicates
    • Compact vertical design saves laboratory bench space

Teknik Özellikler

Analog Outputs 2, 4-20mA (User-configurable concentrations)
Calibration 2 point (KHP two standards)
Certifications CE, EMC EN50082-1, and EN 55011 Group 1 Class A
Data Export To PC via Ethernet, or using a USB memory stick (Microsoft® Excel®-ready .csv file format)
External Dimensions 48.3 cm H x 31.1 cm W x 31.1 cm D (19 in H x 12.25 in W x 12.25 in D)
Gas Requirements < 200 mL/min. 99.99% N2 or CO2-free air
Input Relays 2 (remote start, remote stop)
Instrument Enclosure NEMA 4X / IEC Class IP-56
Measurement Accuracy +5%
Measurement Ranges 0.050 to 25 / 5 to 250 ppm carbon
Measurement Technique Non-dispersive infrared (NDIR) detection
Operating Environment 5 – 45 °C, up to 90% humidity (non-condensing)
Operating Priciple Heated sodium persulfate oxidation
Operator Interface Windows® CE-based, color touchscreen display
Output Relays 2 (system alarm, sample alarm)
Power Requirements 24VDC (Optional 24VDC power supply allows operation with 90-250VAC 50/60Hz source)
Reagents Required Sodium persulfate, phosphoric acid
Regulatory Method Compliance USEPA 415.3 (Source water & drinking water) SM 5310 C (Water & wastewater)
Sample Processing / Analysis Time 4 to 9 minute intervals
Weight 11 kg (24 lbs)